The hourly throughput of the Sysmex XN 1000 is around 100 samples 88 µL in whole blood (WB) mode and body fluid (BF) mode
Reagent management systems up to 100 samples/h
645 (W) x 855 (H) x 755 (D) Approx. 78 Kg
Description
Sysmex XN 1000 is a hematology analyzer that generates a “flag” for individuals who may be at risk of certain hematological disorders based on the CBC results (referred to as an interpretive program (IP) message). Some examples of Sysmex XN-1000 IP messages include RBC agglutination (clumping), iron deficiency, and hemoglobin defects [9].
The accuracy of the Sysmex analyzer to correctly “flag” individuals at high risk of a genetic hemoglobin defect is of critical importance, as this information could be used to refer these high-risk individuals for more definitive testing (which can be expensive and difficult to access).
Product Applications:
XN series enumerate advanced clinical parameters like Ret, He, IPF , NRBC, which are of additional clinical relevance for hospitals with multispeciality functionalities like oncology, nephrology , ICU’s, etc
Ret He, IRF, and reticulocyte count are helpful for anemia management, IG for infections, IPF for screening and monitoring of thrombocytopenic disorders, and NRBC for assessing ineffective erythropoiesis. (RET-He) helps differentiate iron deficiency anemia from anemia of chronic disorders
Features and Benefits of Sysmex XN 1000:
The Sysmex XN 1000 comes with several crucial features, making it a powerful automated analyzer for all types of clinical setups. Following are some essential features and benefits of Sysmex XN 1000:
It comes with a touch screen and graphical user interface, which allows for easy use.
The hourly throughput of the Sysmex XN 1000 is around 100 samples.
Reagent management systems.
You can customize the diagnostic applications based on clinical needs.
It features remote monitoring.
The Sysmex XN 1000 is considerably compact. Hence, it is an ideal option, especially if you have tight laboratory space.
It comes with optional reagent container sizes.
Distinctive features of the Sysmex XN-1000 analyzer:
Choice of diagnostic channels for laboratory tasks:
RET – determination of reticulocytes
PLT-F – fluorescence detection of platelets
WPC – detection of precursors of leukocytes, including stem cells
BF – research of biological fluids
Reflex testing technology (Rerun & Reflex) allows automatic additional examination of samples with an extended profile.
The aspirated sample volume in all modes is 88 μL, which is especially important for pediatrics, neonatology, and gerontology
Capillary blood analysis capability in automatic mode using sample rack AK407696 RBT RACK SET COMPLETE and suitable capillary blood tubes.
Simultaneous work in automatic and manual modes (for emergency samples)
Productivity from 83 to 100 samples per hour depending on the selected mode
Sampler capacity 50 tubes: 5 racks of 10 tubes
Online service access for remote condition monitoring and service for XN analyzers
Cartridge supply of fluorescent dyes. Reagents are easy to change and simple to use
Technical specifications
Measurement principles | WBC DIFF/RET/PLT-F/WPC: Fluorescence Flow Cytometry RBC/PLT: Hydrodynamically focussed impedance HGB: Cyanide-free SLS haemoglobin |
Aspiration volume | 88 µL in whole blood (WB) mode and body fluid (BF) mod |
Throughput | up to 100 samples/h |
Quality control material | XN Check (three levels) and XN Check BF (two levels) |
Parameters | XN-CBC (always with NRBC) & XN-DIFF (always with IG) Whole Blood WBC, NRBC%, NRBC#, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, RDW-SD, RDW-CV, MicroR%, MacroR%, PDW, MPV, P-LCR, PCT, NEUT%, NEUT#, LYMPH%, LYMPH#, MONO%, MONO#, EO%, EO#, BASO%, BASO#, IG%, IG# With RET (optional) RET%, RET#, IRF, LFR, MFR, HFR, RET-He, RBC-He, Delta-He, HYPO-He%, HYPER-He%, PLT-O With PLT-F (optional) PLT-F, IPF, IPF# With WPC (optional) Extended Inflammation Parameters licence: NEUT-GI, NEUT-RI, RE-LYMP%, RE-LYMP#, AS-LYMP%, AS-LYMP# XN Stem Cells licence HPC%, HPC# Body Fluids with XN-BF WBC-BF, RBC-BF, MN%, MN#, PMN%, PMN#, TC-BF# (optional) |
Configurations | XN-1000 [DIFF] XN-1000 [WPC] |