Mindray BC-5150 Price 5 Part Blood Cell Counter
The Mindray BC-5150 Auto 5-Diff Hematology Analyzer is a compact, lightweight 5-part white blood cell differential analyzer designed for small-to-medium clinical laboratories, pediatric clinics, and space-restricted diagnostic centers
Application
Laser Scatter & Flow Cytometry
60 Tests per Hour Speed
Micro Sample Volume 15uL
Complete Blood Count CBC Test
Compact Benchtop Design
Touch Screen TFT Display

Key Specifications & Features
• Test Parameters: 25 reportable parameters, plus research parameters (such as LIC, ALY, and NLR), 3 histograms, and 3 scattergrams.
• Technology: Tri-angle laser scatter, chemical dye, and flow cytometry for WBC 5-part differentiation; DC impedance (50 μm aperture) for RBC and PLT.
• Throughput: Up to 60 samples per hour
Sample Volume:
• Whole blood / Capillary whole blood: 15 μL
• Prediluted mode: 20 μL (high dilution ratio for enhanced mixing)
• Reagents: Uses 3 routine reagents (Diluent, DIFF Lyse, and LH Lyse) with internal placement for select lyse reagents to save bench space.
• Display & Interface: Built-in 10.4-inch TFT color touch screen (eliminating the need for an external PC), 4 USB ports, and a LAN port with bi-directional LIS and HL7 support.
• Data Storage: Up to 40,000+ results with graphics.
• Dimensions & Weight: 400 mm (D) × 320 mm (W) × 410 mm (H); weighs approximately 24 kg

Specifications
5-part differentiation Mindray BC 5150 price
BC-5150
Auto Hematology Analyzer
Technical Specifications
Principles
Impedance method for RBC and PLT counting
Cyanide free reagent for hemoglobin test
Flow Cytometry (FCM) + Tri-angle laser scatter + Chemical dye
Parameters

25 parameters: WBC, Lym%, Mon%, Neu%, Bas%, Eos%, Lym#,
Mon#, Neu#, Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC,
RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC.
4 research parameters including LIC%, LIC#, ALY%, ALY#
3 histograms for WBC, RBC and PLT
Reagent
M-52D Diluent, M-52DIFF Lyse, M-52LH Lyse, Probe Cleanser
Performance
Parameter Linearity Range Precision Carryover
WBC 0-500×10 9 /L ≤2% (4-15×10 9 /L) ≤0.5%
RBC 0-8×1012/L ≤1.5% (3.5-6.0×1012/L) ≤0.5%
HGB 0-250g/L ≤1.5% (110-180g/L) ≤0.6%
PLT 0-5000×10 9 /L ≤6.0% (100-149×10 9 /L) ≤1.0%
≤4.0% (150-500×10 9 /L)
Sample Volume
Prediluted mode 20 μL
Whole blood mode 15 μL
Capillary whole blood mode 15 μL
Throughput
60 samples per hour
Display
10.4 inch TFT Touch Screen

External Thermal printer / Laser printer / Inkjet printer, various
Operating Environment
Temperature: 10°C~30°C
Humidity: 20%~85%
Air pressure:70 kPa~106 kPa
Power requirement
100V-240V
50Hz/60Hz
Dimension and Weight
Depth(400 mm) x width(320 mm) x height(410 mm)
Weight :24kg