CPAP, IMV, triggered modes such as SIMV, SIPPV, Pressure Support Ventilation or Volume Guarantee
olume Guarantee Ventilation: SIMV+VG, SIPPV+VG, PSV+VG
continuous flow, pressure-limited, time-cycled
The Babylog 8000 plus is designed for harmonious ventilation of small children and the smallest pre-term babies. It has an advanced upgrade platform that has allowed it to keep pace with all new forms of treatment and clinical advances. Sensitive synchronisation with gentle but precise support for spontaneous breathing reduces the work of breathing and makes the ventilation process much more comfortable for patients.
The advantages of optimum ventilation
If you use a Babylog 8000 plus, you can rest assured you’re giving the little patient in your care the best possible ventilation. Whatever mode you require
– CPAP, IMV, triggered modes such as SIMV, SIPPV, Pressure Support Ventilation or Volume Guarantee
– the Babylog 8000 plus meets the demands.
The secret of harmonious breathing: milestones in ventilation
For decades there has been a supply of ground-breaking neonatal ventilators. Classic devices that have proved how invaluable they are thousands of times over in locations all around the world. The Babylog 8000 plus is no exception. Its qualities are not confined to functional form, clever design or any other external trappings. The Babylog 8000 plus stands for milestones in neonatal ventilation – inner qualities your little patient will benefit from. The optional Pressure Support and Volume Guarantee functions, for example, bring a new dimension to ventilation therapy and facilitate the process of weaning a baby off the venti lator. Both features contribute to this ventilator’s most valuable asset – its ability to promote harmonious breathing.
The advantages of optimum ventilation: little patients, great demands
If you use a Babylog 8000 plus, you can rest assured you’re giving the little patient in your care the best possible ventilation. Whatever mode you require – CPAP, IMV, triggered modes such as SIMV, SIPPV, Pressure Support Ventilation or Volume Guarantee – the Babylog 8000 plus meets the demands.
TECHNICAL SPECIFICATIONS:
Conventional ventilation: continuous flow, pressure-limited, time-cycled
Triggered ventilation: SIMV, SIPPV, PSV1), leak adapted
- Trigger: Flow/volume trigger, leak adapted
- Trigger delay: approx. 40-60 ms
High Frequency Ventilation: CPAP+HFV, IMV+HFV
- Frequency: 5 to 20 Hz
Volume Guarantee Ventilation: SIMV+VG, SIPPV+VG, PSV+VG
Oxygen mixer loss (bleed flow): 0 (zero) L/min
Settings
Inspiratory oxygen concentration: 21-100 vol.% oxygen
Peak inspiratory pressure: 10 to 80 mbar
PEEP/CPAP: 0 to 25 mbar
Maximal frequency: 200 bpm
Inspiratory time: 0.1 to 2 seconds
Expiratory time: 0.2 to 30 seconds
Inspiratory flow: 1 to 30 L/min
Base Flow (VIVE): 1 to 30 L/min
Monitoring
Flow monitoring: at the Y-piece, integrated
Volume monitoring: at the Y-piece, integrated
Lung function monitoring: compliance, resistance C20/C
Method: linear regression analysis
FiO2 monitoring: integrated
Real-time curves: Flow and pressure, integrated
Inspiratory oxygen concentration: 21-100 vol.% oxygen
Peak pressure: to 99 mbar
Mean airway pressure: to 99 mbar
Graphic trends: 6 parameters, integrated
Logbook: record of up to 100 alarms