The C30 Transport patient monitor is strong, stable, waterproof, dustproof, anti-fall and shockproof, it is Small and lightweight and easy to carry. It can be equipped with a optional specialized bag for first aid monitoring. Suitable for emergency scene, out-of-hospital transport, intra-hospital transport, bedside monitoring in the emergency department
Description
Specifications
Patient: adult, pediatric, neonate
Display: 4.3 “color TFT touch screen
Resolution: 16:9
2 curves on screen
Sweep speed: 12.5mm / s, 25mm / s, 50mm / s
Indicators:
• Battery indicator light
• Dual indicator alarm light
QRS beep and alarm sound
Key sound
AC input socket
MDR
Trend: 48 hours
Alarm: user adjustable high and low limits
Connection to the monitoring center
DVI connection
Breathing
Impedance method
Rate:
• Adult 0-1OOrpm
• Neonatal 0-150 rpm
• Resolution: ± 1 rpm
Apnea alarm: 10s-40s
Audible and visual alarm
Sweep speed: 12.5.25mm/s
Gain selection: x0.25, x0.5, x1, x2
ECG
Cable: 12-lead CardioTecTM,
3 or 5 snaps (selectable)
12 leads: I, Il, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, -V6
5 wire: RA, LA, RL, LL, V or R, L, N, F, C,
3 wire: RA, LA, LL or R, L, F
Lead Selection: 12 leads: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
5 wire: I, Il, IIl, aVR, aVL, aVF, V
3 wire: I, Il, III
Gain selection: x0.25, x0.5, x1, x2, auto
Adult: 15-300 1 / min
Pediatric / neonate: 15-350 1 / min
Accuracy: ± 1% or ± 1 1 / min (whichever is greater)
Resolution: 1 1 / min
ECG bandwidth:
MON: 0.5-55Hz
DIA: 0.05-150 Hz
OPE: 1-20Hz
ST segment detection: ± 2.0mV (automatic)
Arrhythmia Analysis
Pacemaker pulse detection
Temperature
2 channels
Standard sensor: Cutaneous
Measurement range: 0-50 ° C
Resolution: 0.1 ° C
Accuracy: 0.1 ° C (excludes sensor error)
Channel: T1, T2, TD (temperature difference)
PNI
Measurement method: Automatic oscillometric
Working mode: manual / automatic / continuous
Interval: Adjustable from 1 min to 480 min
Measurement types: systolic, diastolic, mean
Unit of measure: mmHg / KPa, adjustable
Systolic pressure range:
• Adult: 40 to 270 mmHg
• Pediatric: 40 to 200mmHg
• Neonate: 40 to 135 mmHg
Diastolic pressure range:
• Adult: 10 to 215 mmHg
• Pediatric: 10 to 150mmHg
• Neonate: 10 to 100 mmHg
Medium pressure range:
• Adult: 20 to 235 mmHg
• Pediatric: 20 to 165mmHg
• Neonate: 20 to 110 mmHg
Overpressure Protection: Hardware and Software
• Accuracy: less than ± 5mmHg
• Resolution: 1mmHg
Sp02 Comen
Range: 1-100%
Resolution: 1%
Accuracy:
• ± 2% (70-100%, Ped, no movement)
• ± 3% (70-100%, Neo, no movement)
• 0-69% unspecified
Range alarm: 0-100%
Pulse Frequency:
• Range: 20 to 300 ppm
• Resolution: 1 ppm
• Accuracy:
– ± 3ppm (no movement)
– ± 5ppm (movement)
Alarm: 20 to 300 ppm
EtCO2
Sidestream CO2 (SideStream)
Range: 0 to 150 mmHg, 0 to 79%,
• 0 to 20 kPa (at 760 mmHg)
• Accuracy: ± 2mm Hg (0-40mmHg)
• ± 5% of reading (41-70 mmHg)
• ± 8% of reading (71-100 mmHg)
• ± 10% of reading (101-150 mmHg)
Frequency: 0-120 rpm
Accuracy:
• ± 2rpm (0-70rpm)
• ± 5 rpm (> 70 rpm)
Response time: <240 ms (10% to 90%)
Delay time: <2s
C02 COMEN Main Flow
Range: 0 to 150 mm Hg, 0 to 79%,
0 to 20 kPa (at 760 mmHg)
Accuracy: ± 2mmHg (0-40mmHg)
• ± 5% of reading (41-70 mmHg)
• ± 8% of reading (71-100 mmHg)
• ± 10% of reading (101-150 mmHg)
Frequency: 0-120 rpm
Accuracy:
• ± 2rpm (0-70rpm)
• ± 5 rpm (> 70 rpm)
Delay time: <2s
Features:
• 4.3-inch 16:9 colorful LCD, high-resolution display, touch screen operation supporting Pingyin, Wubi and handwriting (handwritten graph available)
• Light weight of 1.2kg for easy carrying (carryover graph available)
• Double lithium batteries including a built-in 2600mAh lithium battery of C30 and a built-in 4400mAh lithium battery of a special packet for emergency monitory; above 8 hours of operating time; vehicular power supply supported (battery picture available)
• C30 taken as a module connected with the C90 plug-in type patient monitor (figure of the minihost connected with the major host available )
• Waterproof and shockproof C30 special packet for emergency monitoring capable of holding first aid articles and being carried and used in the field (pictures available)
• 1.2-meter fall protection (picture of fall available)
Standard configuration:
Necllor blood oxygen, electrocardiograph/heart rate, noninvasive blood pressure, blood oxygen, respiration, body temperature, pulse rate
Optional configuration:
end-tidal carbon dioxide, American blood oxygen
Linux Operating System provided
• Simple operating style of Comen's patient monitor, 9 frequently used shortcut keys, suitable for emergency call and first aid (interface picture available)
• Multiple operating interfaces (superposed interface picture available: the interface picture provided by the R&D Department )
• Case history data exchange with the C90 plug-in type patient monitor
• 48-hour trend storage and playback (interface picture provided by the R&D Department available)
• 1-hour short trend display
• 5-minute II electrocardiograph playback interface
• Menu retrospect of 100 ARR alarm events
• Acoustic-optic three-level intelligent alarm, intelligent alarm lamp color switchover between red, yellow and blue, adjustable alarm volume (picture of the alarm lamp available)
First Aid Site (pictures available)
Arrivalat the site of traffic accident (site of accident or site for first aid)
Putting into the ambulance
Lifting onto the litter
Sending into the emergency department
Ward inspection
• Medical personnel can carry the C30 to proceed measurement for patients in wards conveniently
• Medical personnel can also store the measurement information of patients in the C30.
Special Packet for Emergency Monitoring (picture available)
• Waterproof, shockproof and fall protection
• Multiple carrying modes: Single shoulder carrying, litter carrying and ambulance carrying
• Built-in 4400mAh lithium battery
• A left storage bag capable of holding first aid articles and the right storage bag capable of holding all instrument accessories