HACH TL2360 High Range Benchtop Turbidimeter Series
Certifications CE, KC, RCM
Range::

0 – 4000 NTU NTU (Ratio on): 0 – 4000 NTU (Ratio off): 0 – 40 EBC (Ratio on): 0 – 980 EBC (Ratio off): 0 – 9.8

Dimensions::

153 mm x 395 mm x 305 mm 3KG

Price::

$ 8000

PRODUCT DETAILS

Turbidimeter laboratorium Seri terbaru yaitu TL23 memadukan teknologi tepercaya dan fitur yang disempurnakan. Memiliki blok optik yang sama berarti Anda dapat mengganti 2100N / AN lama Anda dengan tubidimeter TL23 tanpa harus mengubah prosedur operasi standar apa pun. Sama seperti instrumen sebelumnya, turbidimeter lab TL23 hadir dalam 4 model berbeda:

• 2100N digantikan oleh TL2300 (EPA, hingga 4.000NTU)

• 2100NIS digantikan oleh TL2310 (ISO, hingga 1.000NTU)

• 2100AN digantikan oleh TL2350 (EPA, hingga 10.000NTU)

• 2100ANIS digantikan oleh TL2360 (ISO, hingga 10.000NTU)

Accuracy:

FNU: ±2% of reading plus 0.01 FNU from 0 - 1000 FN

FAU: ±10% of reading from 20 - 10000 NTU

NTU: ±2% of reading plus 0.01 NTU from 0 - 1000 NTU, ±5% of reading from 1000 - 4000 NTU, ±10% of reading from 4000 - 10000 NTU

Accuracy 3:

Absorbance: ±0.005 Abs from 0 - 1 Abs at 860 nm

Transmittance: 0.12% T from 10 - 100% T at 860 nm

Air Purge:

Dry nitrogen or instrument grade air (ANSI MC 11.1, 1975)

0.1 scfm at 69 kPa (10 psig); 138 kPa (20 psig) max

Hose barb connection for 1/8-inch tubing

Certifications:

CE, KC, RCM

Communication:

USB

Data Logging:

2000 total logs, includes reading log, verification log and calibration log

Dimensions (H x W x D):

153 mm x 395 mm x 305 mm

Display:

17.8 mm (7 in.) color touch screen

Interface:

2 USB-A ports for USB flash drive, external printer, keyboard and barcode scanner

Light Source:

Light-emitting diode (LED) at 860 ± 30 nm

Measurement Method:

Nephelometric

Operating Temperature Range:

0 - 40 °C

Power Requirements (Amps):

3.4 A

Power Requirements (Hz):

50/60 Hz

Power Requirements (Voltage):

100 - 240 VAC

Range:

FNU (Ratio on): 0 - 1000

FNU (Ratio off): 0 - 40

FAU (auto range): 20 - 10000

NTU (Ratio on): 0 - 10000 auto decimal

NTU (Ratio off): 0 - 40

EBC (Ratio on): 0 - 2450 auto decimal

EBC (Ratio off): 0 - 9.8

Absorbance (auto range): 0 - 2.00

Transmittance (%): 1.0 - 100

Degree (mg/L): 0 - 100

Reading Modes:

Manual or auto range, signal averaging on and adjustable or off, ratio on or off

Regulatory:

Meets ISO 7027, DIN EN 27027, DIN 38404 and NFT 9033

Repeatability:

±1% of reading or 0.01 FNU/NTU, whichever is greater (under reference conditions)

Response Time:

Signal averaging off: 6.8 seconds / Signal averaging on: 14 seconds (when 10 measurements are used to calculate the average)

Sample Cell Compatibility:

Round cells 95 x 25 mm (3.74 x 1 in.) borosilicate glass with rubber-lined screw caps


Note: Smaller sample cells (less than 25 mm) can be used when a cell adapter is used.

Sample Requirements:

25 mm sample cell: 20 mL minimum

0 to 70 °C (32 to 158 °F)

Stabilization Time:

Immediately

Storage Conditions:

-20 to 60 °C

Units:

FNU, FAU, NTU, EBC, Abs (absorbance), %T (% transmittance) and mg/L

Weight:

3.0 kg