IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
8.4" TFT color LCD screen display
2 years
AV-2000B2 Multi-functional Electric Hospital Vertical ICU Ventilator AV-2000B2 Multi-functional Electric Hospital Vertical ICU Ventilator
AV-2000B2 is a multi-functional Ventilator, adopting pneumatic, electronic control and microcomputer technology. Imported components are utilized to guarantee stablility and reliabiligyt. With color screen(TEF) to display various measuringand configuration parameters, the ventilator suitable for respiratory treatment, applicable for medical,surgical and gynecological department, pediatrics department, first aid, ICU etc.Air source cen be conneted.
Product Introduction:
AV-2000B2 is a multi-functional Ventilator, adopting pneumatic, electronic control and microcomputer technology. Imported components are utilized to guarantee stablility and reliabiligyt. With color screen(TEF) to display various measuringand configuration parameters, the ventilator suitable for respiratory treatment, applicable for medical,surgical and gynecological department, pediatrics department, first aid, ICU etc.Air source cen be conneted.
Ventilator:
Display mode | High-definition 8.4” TFT color LCD screen display |
Ventilation mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH |
Ventilation parameters | Tidal volume : 50~1500 ml Rate : 2~99 bpm SIMV rate : 2~20 bpm I:E : 2:1~1:8 Inspiratory trigger pressure: -1.0~2.0kPa PEEP: 0~2.0 kPa Pressure range : 1.0~6.0 kPa SIGH : 1.5 times the inspiratory time Inspiratory platform : 0~50% |
Parameters for ventilation monitoring | tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform |
Monitoring of oxygen concentration | 21%~100% |
Safety alarm system
Oxygen Concentration alarm | upper limit setting range 50%~100% low limit setting range 15%~50% |
Airway pressure alarm | upper limit setting range 1.0~6.0 kPa low limit setting range 0.4~2.0 kPa |
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min low limit setting range 1.0~10L/min |
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds |
Power | AC 220V±10% 50Hz±1Hz UPS, Storage battery |